Flir T600-Series

 

 

FLIR T600

FLIR T620

FLIR T640

Imaging performance

Resolution

480x360 pixels

640x480 pixels

640x480 pixels

Spatial resolution

0.91 mrad for 25° lens
0.55 mrad for 15° lens
1.64 mrad for 45° lens

0.68 mrad for 25° lens
0.41 mrad for 15° lens
1.23 mrad for 45° lens

0.68 mrad for 25° lens
0.41 mrad for 15° lens
1.23 mrad for 45° lens

Thermal sensitivity (at 30 °C)

<40 mK @ 30 °C

<40 mK @ 30 °C

<35 mK @ 30 °C

Digital zoom

1-4x continuous, incl. panning

1-4x continuous, incl. panning

1-8x continuous, incl. panning

Focus

Automatic (one shot) or manual

Automatic (one shot) or manual

Continuous, one shot or manual

 

 

 

 

Image presentation

MSX

N/A

N/A

IR image with MSX

Viewfinder

N/A

N/A

800x480 pixels

 

 

 

 

Image annotation

Image sketch

N/A

N/A

On IR and visual image

 

 

 

 

Report generation

Instant Report in camera

N/A

Automatic generation of PDF report based on selected images direct in camera

Automatic generation of PDF report based on selected images direct in camera

 

 

 

 

Measurement

Temperature range, standard

-40 °C to +150 °C
+100 °C to +650 °C

-40 °C to +150 °C
+100 °C to +650 °C

-40 °C to +150 °C
+100 °C to +650 °C
+300 °C to +2,000 °C

Temperature range, optional

+300 °C to +2,000 °C

+300 °C to +2,000 °C

 

 

 

 

 

Measurement analysis

Line profile function

N/A

N/A

Live profile, H/V-direction

Measurement presets

N/A

N/A

Yes

 

 

 

 

Geographic Information System

Built-in GPS

N/A

Location data automatically added to every image for referencing on WEB maps

Location data automatically added to every image for referencing on WEB maps

General specifications

 

Imaging performance

Field of View (FOV) / minimum focus distance

25° x 19° / 0.25 m
15° x 11° / 0.5 m
45° x 34° / 0.15 m
lens needs to be specified when ordering

Focal Plane array (FPA)

Uncooled microbolometer 640x480 pixels, latest generaton with 17 µm pitch

Spectral range

7.8 to 14 µm

Image frequency

30 Hz

Focus

Manual / autofocus

 

 

Image presentation

Display

4.3" superbright touchscreen LCD 800x480 pixels

Image modes

IR-image with selected color scale, Full color visual, Picture in Picture (Resizable and movable IR-area), Thermal Fusion (Threshold above, below and interval), thumbnail gallery

Manual image adjustments

Level/span/max/min

Automatic image adjustments, continuous or manual activation

Standard or based on histogram from image content

Automatic image adjustment with locked scale

Lock max, min or span

 

 

Measurement

Accuracy

± 2 °C or ± 2% of reading

 

 

Measurement analysis

Spotmeter

10

Area

5 Max/Min/Average value within box or circle

Automatic hot/cold detection

Max/Min temp. value and position shown within box, circle or on a line

Isotherm

Detect high/low temperature/interval

Difference temperature

Difference between any two measurement functions or any measurement function and a reference temperature

Reference temperature function

Manually set

Emissivity correction

Variable from 0.01 to 1.0 or selected from materials list

Measurement corrections

Reflected temperature, optics transmission and atmospheric transmission

External windows correction

Automatic based on inputs of window temperature and transmission

Measurement function alarm

Audible/visual alarms (above/below) on any selected measurement function

 

 

Set-up

Image controls

Palettes (Arctic, Gray, Iron, Lava, Rainbow and Rainbow HC), image adjustment (auto/manual)

Set-up controls

Local adaptation of units, language, date and time formats; automatic shutdown , display intensity

Configure information to be shown in image

Datasheet